5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-291
SMPL>M Page: O, S Stand: M, A Abbreviation: SMPL>
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SMPL>M ..... V.fraction ........mL V.total .........L
Function Add sample volume to the measuring vessel (MEAS).
Parameter Range; Default value / Meaning
V.fraction 0.010...50.000 mL; -
Fractional volume
With diluted samples, here the volume fraction of the total volume
V.total which is added to the sample vessel must be entered. The
following condition holds:
V.fraction V.total
V.total 0.10 mL...100000 L; -
Total volume
With diluted samples, here the total volume of sample and dilution
solution must be entered. The following condition holds:
V.total V.fraction
Description The SMPL>M instruction is used for the manual or automatic addition of
the sample volume to the measuring vessel (MEAS) on the 747 VA
Stand:
In manual addition or addition with the 6.5611.010 Pipet-
ting equipment, the prompt SMPL "V.sample" > M for
manual addition of the sample volume V.sample appears
in the program run.
In automatic addition with the 695 Autosampler, the sam-
ple volume is transferred from the sample vessel on the
sample rack defined on the 1
st
sample data line to the
measuring vessel.
With undiluted samples (V.fraction and V.total blank), the sample
volume is given either by S0 (on entry of S0 as a volume) or by S0/S2
(on entry of S0 as a weight, S2 is the density in g/mL).
With diluted samples (V.fraction > 0 and V.total V.fraction), the
sample volume is given by V.fraction, irrespective of the entry of S0.
You will find further details on the sample size in the description of the
parameter Sample size/S0 in section 5.4.4.
see also SMPL/M, SMPL/F