Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-289
SMPL/F Page: O, S Stand: A Abbreviation: SMP
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SMPL/F ..... V.fraction ........mL V.total .........L
Function Sample volume has been added to the sample vessel (FEED) on the 695
Autosampler.
Parameter Range; Default value / Meaning
V.fraction 0.010...50.000 mL; -
Fractional volume
With diluted samples, here the volume fraction of the total volume
V.total which has been added to the sample vessel must be entered.
The following condition holds:
V.fraction ≤≤ V.total
V.total 0.10 mL...100000 L; -
Total volume
With diluted samples, here the total volume of sample and dilution
solution must be entered. The following condition holds:
V.total ≥≥ V.fraction
Description The SMPL/F instruction defines the sample volume in the sample vessel
(FEED) on the 695 Autosampler.
With undiluted samples (V.fraction and V.total blank), the sample
volume is given either by S0 (on entry of S0 as a volume) or by S0/S2
(on entry of S0 as a weight, S2 is the density in g/mL).
With diluted samples (V.fraction > 0 and V.total ≥≥ V.fraction), the
sample volume is given by V.fraction, irrespective of the entry of S0.
You will find further details on the sample size in the description of the
parameter Sample size/S0 in section 5.4.4.
Comments • The SMPL/F instruction can be used only in operation with the 695
Autosampler.
• If no other dispensing operations are performed in the sample vessel
(FEED), an asterisk * can also be entered for Sample size/S0. The
sample volume is then defined by the subsequent transfer into the
auxiliary or measuring vessel.
see also SMPL>M, SMPL/M