5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-286
SIMRESET Page: O, S Stand: M, A Abbreviation: SIMR
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SIMRESET .....
Function Resetting the 3.496.8380 VA Simulator.
Description The SIMRESET instruction is used to reset the current increases initiated
on the 3.496.8380 VA Simulator with the SIMPULSE instruction for the
simulation of standard additions.
Comments The 3.496.8380 VA Simulator and the associated 3.496.5190
Connecting cable are available as an option. The VA Simulator gen-
erates 2 polarographic current waves with half-wave potentials of ca.
–540 mV and ca. –1070 mV in a defined voltage range. You will find
further details in the 3.496.8380-SB Circuit Description supplied with
the VA Simulator.
To ensure that the SIMRESET instruction functions properly, a line time
1 s must be entered.
Example see SIMPULSE
see also SIMPULSE