Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-283
SETCTRL Page: O, S Stand: M, A Abbreviation: SET
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SETCTRL ..... Code ........ Message ................
Function Set the output control lines 1...8.
Parameter Range; Default value / Meaning
Code 1, 0, *; *
Code for output control lines
8-digit numeric code for setting the output control lines 1...8. The
following values can be entered for each line:
0 off, inactive (open)
1 on, active (0 V)
* leave status unchanged
Message 16 characters;
Message for output control lines
Possibility for entry of a remark concerning the command.
If this field is edited, 4 preset messages for operation of the 708 Sam-
pling Unit are displayed on the softkeys. If such a message is entered
by pressing one of the softkeys, the last three digits of the code are
simultaneously changed:
FILL PUMP ON ÞÞ Code = *****101 Injection valve to "FILL",
pump on
FILL PUMP OFF ÞÞ Code = *****100 Injection valve to "FILL",
pump off
INJECT PUMP ON ÞÞ Code = *****011 Injection valve to "INJECT",
pump on
INJECT PUMP OFF ÞÞ Code = *****010 Injection valve to "INJECT",
pump off
Description The SETCTRL instruction is used to set the output control lines 1...8 of the
remote interface "Control Lines" to the status preset in Code. Functions
can thus be initiated at external devices attached to the remote interface.
Comments The attachment of devices to the remote interface is described in section
3.13.
Example see ADD/M
see also SCANCTRL