Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-278
SDOS>F Page: O, S Stand: A Abbreviation: SDOS>F
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SDOS>F ..... Soln.name ........ Prop.factor ........
Function Dispense auxiliary solution proportional to the sample size S0 into the
sample vessel (FEED) on the 695 Autosampler.
Parameter Range; Default value / Meaning
Soln.name 8 characters; -
Name of the auxiliary solution
The entry of a solution name is mandatory when the auxiliary solution
should be added automatically. In this case the solution name
Soln.name must be entered in the parameter of the same name on the
"SOLUTIONS" page and assigned a position Pos. (see section 5.3.4).
Without entry of a name for Soln.name in the SDOS>F instruction,
without entry of this name on the "SOLUTIONS" page or without a
position assignment, the auxiliary solution must be dispensed manually
(the prompt DOS "Soln.name" "V.add" > F then appears during the
program run).
Prop.factor 0.0010...99999.9; 1
Proportionality factor
Specification of the proportionality factor for calculation of the addition
volume V.add:
V.add (mL) = Prop.factor * S0 (mL or g)
Description The SDOS>F instruction is used to dispense an auxiliary solution (e.g.
supporting electrolyte, reagent solution) proportional to the sample size
S0 into the sample vessel (FEED) on the 695 Autosampler.
For automatic addition with the 695 Autosampler, the appropriate
position of the auxiliary solution must be specified on the "SOLUTIONS"
page (see Soln.name).
Comments The SDOS>F instruction can be used only in operation with the 695
Autosampler.
see also SDOS>A, SDOS>F, DOS/M, DOS>M, DOS>F, DOS>A, PDOS>M