Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-274
REP) Page: O, S Stand: M, A Abbreviation: REP
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. REP)n .....
Function End of the loop for replications.
Parameter Range; Default value / Meaning
n 0...8; 0
Number of replications
This parameter defines the number of times the REP loop is repeated.
Description The REP)n instruction ends a loop opened with (REP for replications.
The instructions inserted in this loop are repeated n times.
Comments see (REP
Example see (REP
see also (REP, (ADD, ADD)n, (CAL, CAL)n, (VAR, VAR)n