Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-270
Description The RAMP instruction starts the voltage ramp defined in the parameters.
U
t
U.start
U.end
U.step
t.step
Comments The electrode type must be defined before the RAMP instruction.
Admissible are HMDE or RDE.
The measurement mode must be defined before the RAMP instruc-
tion. Admissible is only CYCMODE.
The RAMP instruction must always be located within a CSWEEP
block. Maximum 5 RAMP instructions are allowed.
Apart from the RAMP instructions, no other SWEEP instruction is
allowed in the same segment.
If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
Within a CSWEEP block a wait time between two ramps can be
programmed only by insertion of an additional, slow ramp as flat as
possible.
Examples Cyclic voltammetry of Pb and Cd at the HMDE:
Method: cv SEGMENT
CdPb
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 HMDE Drop size 2 Meas.cell normal
2 CYCMODE I.range 1.4 uA Prep.cycles 2
Meas.cycles 5
3 (CSWEEP
4 RAMP U.start -100 mV U.step 10 mV
U.end -900 mV t.step 0.10 s
Sweep rate 100 mV/s
5 RAMP U.start -900 mV U.step 10 mV
U.end -100 mV t.step 0.10 s
Sweep rate 100 mV/s
6 CSWEEP)
7 END .
see also CYCMODE, (CSWEEP, CSWEEP), HMDE, RDE