Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-253
DOS>M instruction, the addition volume V.add can not be preallocated,
but is recalculated in every measurement from the evaluation value
EV:<subst> measured beforehand for any substance and a proportion-
ality coefficient Prop.coeff..
The automatic addition can be performed with a 685 Dosimat, a 700
Dosino or directly with the 695 Autosampler. The appropriate position of
the standard addition solution must be specified on the "SOLUTIONS"
page (see Soln.name).
Comments On execution of the PDOS>M instruction with a 685 Dosimat or 700
Dosino, a wait is made until the addition is complete before the
instruction following PDOS>M is executed.
see also DOS/M, DOS>M, DOS>A, DOS>F, SDOS>M, SDOS>A, SDOS>F