Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-241
Description The FSWEEP instruction (Fast Sweep) starts the voltage sweep defined in
the parameters. During the sweep, the potentiostat always operates in
the operating mode "fast".
U
t
U.start
U.end
U.step
t.step
Comments • The electrode type must be defined before the FSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
• The measurement mode must be defined before the FSWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
• The use of the FSWEEP instruction is practical only with the measure-
ment modes DPMODE and SQWMODE. With the other allowable meas-
urement modes, FSWEEP is identical to SWEEP.
• The FSWEEP instruction should always be used instead of the SWEEP
instruction with adsorptive stripping techniques (e.g. Ni and Co with
DMG).
• With the FSWEEP instruction maximum 1000 measured values can be
recorded.
• Apart from the FSWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
• If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
see also SWEEP, DSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DPMODE,
SQWMODE