Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-241
Description The FSWEEP instruction (Fast Sweep) starts the voltage sweep defined in
the parameters. During the sweep, the potentiostat always operates in
the operating mode "fast".
U
t
U.start
U.end
U.step
t.step
Comments The electrode type must be defined before the FSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
The measurement mode must be defined before the FSWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
The use of the FSWEEP instruction is practical only with the measure-
ment modes DPMODE and SQWMODE. With the other allowable meas-
urement modes, FSWEEP is identical to SWEEP.
The FSWEEP instruction should always be used instead of the SWEEP
instruction with adsorptive stripping techniques (e.g. Ni and Co with
DMG).
With the FSWEEP instruction maximum 1000 measured values can be
recorded.
Apart from the FSWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
see also SWEEP, DSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DPMODE,
SQWMODE