Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-237
F>A Page: O, S Stand: A Abbreviation: F
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. F>A ..... addition FEED -> AUX V.add ........mL
Function Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.
Parameter Range; Default value / Meaning
V.add 0.010...50.000 mL; 1.000 mL
Transfer volume
Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.
Description The F>A instruction transfers the solution volume specified under V.add
from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.
Comments The F>A instruction can be used only in operation with the 695
Autosampler.
The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).
If the solution volume in the sample vessel < V.add, an error
message appears.
see also A>M, F>M