Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-237
F>A Page: O, S Stand: A Abbreviation: F
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. F>A ..... addition FEED -> AUX V.add ........mL
Function Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.
Parameter Range; Default value / Meaning
V.add 0.010...50.000 mL; 1.000 mL
Transfer volume
Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.
Description The F>A instruction transfers the solution volume specified under V.add
from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.
Comments • The F>A instruction can be used only in operation with the 695
Autosampler.
• The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).
• If the solution volume in the sample vessel < V.add, an error
message appears.
see also A>M, F>M