5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-235
EXIT Page: O, S Stand: M, A Abbreviation: EX
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. EXIT .....
Function Immediate exit from a loop or a block if the test condition is met.
Parameter Range; Default value / Meaning
(Test) EV:<subst>,MC:<subst>,S0,S1,S2,S3,V0 <,<=,>=,> const.,C##;
STATE =,<> ctrl.line code; -
Test condition for decision
The following test conditions can be entered in the field under Main
parameters:
EV:<subst> < | <= | >= | > const. / C##
MC:<subst> < | <= | >= | > const. / C##
S0 < | <= | >= | > const. / C##
S1 < | <= | >= | > const. / C##
S2 < | <= | >= | > const. / C##
S3 < | <= | >= | > const. / C##
V0 < | <= | >= | > const. / C##
STATE = | <> ctrl.line code
The terms have the following meaning:
EV:<subst> Evaluation value (I.peak, I.wave, P.peak or
t.peak) for the specified substance (<subst>
= substance name) in A, W or s
MC:<subst> Mass concentration ρ
s
for the specified sub-
stance (<subst>=substance name) in g/L
S0 Sample size S0 in g or L
S1,S2,S3 Sample specific variables S1, S2, S3
V0 Sample volume in measuring vessel
V0.sample in L
STATE State of the input control lines 1...8 of the
remote interface
const. Constant (numeric value)
C## Common variable C01 ... C18 from the
working storage (see section 5.4.5)
ctrl.line code 8-digit numeric code for the state of the in-
put control lines 1...8. The following values
can be entered for each line:
0 off, inactive (5 V)
1 on, active (0 V)
* immaterial