Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-234
END Page: O, S Stand: M, A Abbreviation: E
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## END .....
Function End of the main program ("OP.SEQUENCE") or end of the segment
("SEGMENTS").
Description The END instruction closes the main program on the "OP.SEQUENCE"
page or a segment on the "SEGMENTS" page.
Comments Each program or segment must be closed with the END instruction.
The END instruction in the main program switches off all functions still
active at the 747 VA Stand (e.g. stirrer, purging, electrodes, voltage)
with exception of the standby voltage U.standby defined in a ØMEAS
instruction.
In contrast to END in the main program, the END instruction in the
segment simply causes a return to the location in the main program
where the segment was called up.
see also STIR, PURGE, DME, HMDE, SMDE, RDE, MEAS