Description The DSWEEP instruction (Dummy Sweep) starts the voltage sweep
defined in the parameters. While measured values are recorded during
the sweep, they are not stored.
U
t
U.start
U.end
U.step
t.step
Comments • The electrode type must be defined before the DSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
• The measurement mode must be defined before the DSWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
• With the DSWEEP instruction, the potentiostat always operates in the
operating mode "fast".
• If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
see also SWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DCTMODE,
DPMODE, AC1MODE, AC2MODE, SQWMODE