Description The DSWEEP instruction (Dummy Sweep) starts the voltage sweep
defined in the parameters. While measured values are recorded during
the sweep, they are not stored.
U
t
U.start
U.end
U.step
t.step
Comments The electrode type must be defined before the DSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
The measurement mode must be defined before the DSWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
With the DSWEEP instruction, the potentiostat always operates in the
operating mode "fast".
If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
see also SWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DCTMODE,
DPMODE, AC1MODE, AC2MODE, SQWMODE