5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-231
Adjustable parameters:
DPMODE
t.step
t.meas
t.pulse
U.ampl
SWEEP
U.start
U.end
U.step
F (Hg)
t.drop
t
U
t
U.step
U.start
I
U
t.meas (
I
1
)
U.peak
I.peak
U.ampl
t.stept.pulse
t.meas (
I
2
) t.meas (
I
1
) t.meas (
I
2
)
I = I
2
- I
1
Comments The DPMODE measurement mode can be used with all types of
electrodes together with the instructions MEAS, SWEEP, DSWEEP,
FSWEEP or TSWEEP.
The electrode type DME, SMDE, HMDE or RDE must be entered before
the DPMODE measurement mode.
If the DPMODE measurement mode is used with the voltage sweep
SWEEP and the electrode type HMDE, the potentiostat normally oper-
ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the pulse time t.pulse < 30.05 ms.
If the DPMODE measurement mode is used with SWEEP and the
electrode type SMDE or DME, the setting of the potentiostat depends
on the measurement profile. Normally, a switch is also made from
"slow" to "fast" when the pulse time t.pulse < 30.05 ms. With very
small signals, however, the operating mode "slow" is used and with
very large signals the operating mode "fast", irrespective of the pulse
time.
If the DPMODE measurement mode is used with the voltage sweep
SWEEP and the electrode type RDE, the potentiostat always operates in
the operating mode "fast".
If the DPMODE measurement mode is used with the voltage sweep
FSWEEP, the potentiostat always operates in the operating mode "fast".
This mode is recommended especially for adsorptive stripping tech-
niques (e.g. Ni and Co with DMG).
The quantity t.drop important for the definition of t.step depends
on the type of electrode selected:
DME t.drop = 20 ms
HMDE and RDE t.drop = 0 ms
SMDE t.drop = Drop.size ×× 40 ms
see also SWEEP, FSWEEP, DME, HMDE, SMDE, RDE
Area (Hg)