Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-222
The current curves obtained with the CYCMODE measurement mode
can normally not be evaluated. An evaluation is possible only when
just 1 ramp is used and a peak-shaped current profile is found (the
evaluation then corresponds to the case DCTMODE with SSWEEP, i.e.
waves can not be evaluated).
Adjustable parameters:
CYCMODE
I.range
Prep.cycles
Meas.cycles
RAMP
U.start
U.end
U.step
t.step
U.ox
U.start
(ramp 1)
t.step
I
cath
U
I.range
Prep.cycles Meas.cycles
U.step
U.end
(ramp 1)
U.end
(ramp 2)
U.start
(ramp 2)
I
an
U.red
t.HMDE
t
U
t
F (Hg)
see also (CSWEEP, CSWEEP), RAMP, HMDE, RDE
An evaluation is
possible only for
single ramps
For every measure-
ment cycle, maximum
5 ramps are possible
Area (Hg)