Owner manual
3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-10
7
Install MME or dummy stopper
in operation with MME:
• Carefully insert multi-mode electrode 3838 (6.1246.020) in opening 7676 (the
underside of the capillary must not touch the measuring head during in-
sertion) and push in as far as it will go.
• Screw FEP tubing 5151 (6.1805.180) for inert gas supply into connection 9393
of the MME 3838.
• Screw FEP tubing 5959 (6.1805.180) for inert gas supply into connection 9494
of the MME 3838.
• Attach electrode cable 3737 (WE) to screw connection 110110 of the MME 3838:
push cable lug under the screw and then tighten screw firmly.
in operation with RDE (option):
• Insert stopper 119119 (6.2709.040, option) into opening 7676 as far as it will go
so that the two blind holes point to the rear of the stand.
• Screw FEP tubing 5151 (6.1805.180) into upper hole of stopper 119119.
• Screw FEP tubing 5959 (6.1805.180) into lower hole of stopper 119119.
8
Install accessories for rinsing and siphoning off (option)
For rinsing the measuring vessel at the 747 VA Stand manually (dialog
page "STAND") or with the RINSE command, two 700 Dosinos with
50 mL Dosing units must be connected to the 746 VA Trace Analyzer
and initialized acc. to section 3.10.5 and the 6.1456.210 measuring
vessel (option) must be used instead of the 6.1415.210 measuring
vessel.
• Cut PTFE tube 9191 (6.1819.010) to a length of ca. 30 mm and insert from
above in opening 7474 of measuring head 4141 (see Fig. 10).
• Cut the bottom end of PTFE tube 9292 (6.1819.010) diagonally and insert
from above in opening 7373 of measuring head 4141. To ensure that the solu-
tion is siphoned off as completely as possible, the end of the tube must
be located in the deepest part of the 6.1456.210 measuring vessel (left
rear when viewed from front).
• Screw FEP tubing 8989 (6.1805.180) into the threaded openings 7373 and 8787.
• Screw FEP tubing 9090 (6.1805.100) into threaded opening 7474 of measur-
ing head 4141. Screw a 6.1808.000 tubing coupling to the other end of FEP
tubing 9090 and insert the coupling in a slot of the tubing holder at the rear
of the VA Stand.
9
Install measuring vessel
• Tilt back measuring head arm 4040.
• Slide measuring vessel 2828 into holder 2626 from the front and fill with
analyte solution or dist. H
2
O (storage solution) until the tips of the MME
and the reference electrode are immersed in the liquid.
• Lower measuring head arm 4040 and cover 2424.