Owner manual
5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-204
CAL)n Page: O, S Stand: M, A Abbreviation: C
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. CAL)n .....
Function End of the loop for recording the calibration curve.
Parameter Range; Default value / Meaning
n 0...9; 0
Number of calibration solutions – 1
This parameter defines the number of times the CAL loop is executed,
i.e. how many calibration solutions are used to determine the calibra-
tion curve.
Description The CAL)n instruction ends a loop opened with (CAL for the recording
of a calibration curve with the calibration technique with calibration curve
cal.crv. or with internal standard int.std..
In calibrations with a calibration curve, for the recording of this curve
the CAL loop must include one of the dosing instructions CSOLN/M,
CSOLN>M or CSOLN/F for addition of the calibration solutions as well as a
sweep or segment call-up. In addition, any other instructions can be
incorporated.
In calibrations with an internal standard, for the recording of the
calibration curve to determine the proportionality factor the CAL loop
must include one of the dosing instructions CSOLN/M, CSOLN>M or
CSOLN/F for addition of the calibration solutions, the instruction RADD>M
or RADD>F for addition of the internal standard as well as a sweep or
segment call-up. In addition, any other instructions can be incorporated.
Comments see (CAL
Examples
see (CAL
see also (CAL, CSOLN/M, CSOLN>M, CSOLN/F, EXIT,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)