Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-199
BACKGND) Page: O Stand: M, A Abbreviation: BA
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. BACKGND) .....
Function End of the block for background compensation.
Application The background compensation is used to compensate interference of
the measured curves due to the supporting electrolyte. Such interference
includes both the presence of the analyte (blank value) and that of
foreign substances electroactive in the same range.
Description The BACKGND) instruction closes a block opened with the (BACKGND
instruction for the recording of a measured curve for the background
compensation, which is then subtracted from all subsequent measured
curves of the same segment.
The BACKGND block must include a segment call-up, in addition other
instructions whatever their nature can be incorporated.
Comments see (BACKGND
Example see (BACKGND
see also (BACKGND