Owner manual
5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-193
Example Addition of a standard addition solution with the 665 Dosimat via the
remote interface:
746 connection: Connection "Control Lines" 2323 (see section 3.13)
665 connection: Connection 88 for external dosing contact
(see 665 Instructions for Use)
Connecting cable: Cable with 25-pin connector, pin 5 (output 1) connected to red
banana plug, pin 25 (0 V) to black banana plug
Setting on 665: "DIS R" mode with fixed dispensing volume V.add
Program:
:
SMPL>M or SMPL/M Sample addition to measuring vessel
(ADD Loop start for standard additions
(REP Loop start for replication
SEGMENT Call-up of the segment (measurement)
REP)n Loop end, n = number of replications
SETCTRL *******1 Start of dispensing at 665 Dosimat
: by setting control line 1 active
NOP xxx s Entry of a wait time until dispensing
: is at an end
SETCTRL *******0 Resetting of control line 1
ADD/M Standard addition solution has been added
ADD)n Loop end, n = number of standard additions
:
see also (ADD, ADD)n, ADD>M, PADD>M, SETCTRL, SEND>RS#