Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-191
ADD)n Page: O, S Stand: M, A Abbreviation: AD
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. ADD)n .....
Function End of the loop for standard addition.
Parameter Range; Default value / Meaning
n 0...8; 0
Number of standard additions
This parameter defines the number of repetitions of the ADD loop, i.e.
the number of standard additions.
Description The ADD)n instruction ends a loop opened with (ADD for standard
additions. This loop must include one of the dispensing instructions
ADD/M, ADD>M or PADD>M for addition of the standard addition solution as
well as call-up of a sweep or segment. Any other additional instructions
can be incorporated.
Comments see (ADD
Example
see (ADD
see also (ADD, ADD/M, ADD>M, PADD>M