5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-186
The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type
linear, polynom or expon. (see section 5.6.3).
I
U
U.peak
I.peak
Comments The AC1MODE measurement mode can be used with all types of
electrodes together with the instructions MEAS, SWEEP, DSWEEP,
FSWEEP or TSWEEP.
The electrode type DME, SMDE, HMDE or RDE must be entered before
the AC1MODE measurement mode.
With the AC1MODE measurement mode, the potentiostat always
operates in the operating mode "fast". SWEEP and FSWEEP are thus
identical for this mode.
The quantity t.drop important for the definition of t.step depends
on the type of electrode selected:
DME t.drop = 20 ms
HMDE and RDE t.drop = 0 ms
SMDE t.drop = Drop.size ×× 40 ms
see also AC2MODE, SWEEP, DME, SMDE, HMDE, RDE
n = Meas.cycles
Σ(I
n+
– I
n–
)
n
I =