Owner manual
5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-186
The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type
linear, polynom or expon. (see section 5.6.3).
I
U
U.peak
I.peak
Comments • The AC1MODE measurement mode can be used with all types of
electrodes together with the instructions MEAS, SWEEP, DSWEEP,
FSWEEP or TSWEEP.
• The electrode type DME, SMDE, HMDE or RDE must be entered before
the AC1MODE measurement mode.
• With the AC1MODE measurement mode, the potentiostat always
operates in the operating mode "fast". SWEEP and FSWEEP are thus
identical for this mode.
• The quantity t.drop important for the definition of t.step depends
on the type of electrode selected:
DME t.drop = 20 ms
HMDE and RDE t.drop = 0 ms
SMDE t.drop = Drop.size ×× 40 ms
see also AC2MODE, SWEEP, DME, SMDE, HMDE, RDE
n = Meas.cycles
Σ(I
n+
– I
n–
)
n
I =