Owner manual

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-172
Parameter Range; Default value / Meaning
Curve Orig, Curve Proc, Curve Smth, Curve Derv, Curve PSA
For the output of these curves, entry of the substance or segment
name and the VR number is necessary:
Sbst:<subst> Output of substance curve.
The scalings of the axes are in accordance with
the display/plot parameters for the substances
(see section 5.6.3).
In measurements with background compensa-
tion, the background is already subtracted.
At the end of the curve output, the substances
found with name, peak/wave voltage (U) and
evaluation quantity current (I), power (P) or time (t)
are put out.
Sgmt:<segm> Output of segment curve.
The scalings of the axes are in accordance with
the display/plot parameters for the segments (see
section 5.6.2).
In measurements with background compensa-
tion, the curve 00-backgnd is also put out in addi-
tion to the 00 curve if VR = 00 or VR = ** is
selected.
VR: xx Selection of the single measurement whose curve
should be plotted by specifying the VR number (V
= variation, R = replication). On entry of *, all
variations or replications are put out.
Curve Cal
For output of the calibration curve, entry of the substance name is
necessary:
Sbst:<subst> Output of the substance calibration curve. The
scaling of the y axis (EV) is automatic, that of the x
axis (ρ) in accordance with the parameter
M.conc./cm entered on the "SUBSTANCES" page
(see section 5.6.3).
The name of the selected substance is put out following the calibra-
tion curve.
TO destination Printer, RSIfc.1, RSIfc.2; Printer
Destination of automatic output
The automatic output of reports or curves can be sent to the following
destinations:
Printer Printer connected to parallel interface (746.0010)
or built-in thermal printer (746.0020)
RSIfc.1 RS232 interface 1
(attachment possibilities, see section 3.12)
RSIfc.2 RS232 interface 2
(attachment possibilities, see section 3.12)
If Printer has been selected as the first destination, one of the two RS
interfaces can also be selected as a second destination for simulta-
neous output.