Owner manual
5.6.5 DOCUMENTATION
746 VA Trace Analyzer / 747 VA Stand
5-169
Parameter Range; Default value / Meaning
Report Comment Comment report with output of the method com-
ment entered on the subpage "Comment".
Report ActDetm Full report of the determination including determi-
nation file for output to a PC via RS interface for
further processing with the PC Windows pro-
grams "VA Back Up" and "VA Database". The re-
port contains an ASCII part and finally the binary
file. It has the following layout:
XXXXXXXX.dtm File name of determination
A-Fullreport, esize: Identification for full report and
details of file size in bytes
Report Full Full report (see above)
A-Meth.spec Identification for method report
Report MethSpc Method report (see above)
A-Sweeps (smooth) Identification for report of
smoothed measured points
Report Mpts List of smoothed measured
points for all segments (see
above)
A-Baselines Identification for report of base-
lines
Report Mpts List of baseline points for all sub-
stances
A-Calibrations Identification for report of calibra-
tion curves
Report Mpts List of calibration points for all
substances
B-File Identification for binary file
XXXXXXXX.dtm Determination file (binary file)
Report ActMeth Full method report including method file for output
to a PC via RS interface for further processing
with the PC Windows programs "VA Back Up" and
"VA Database". The report includes an ASCII part
and finally the binary file. It has the following lay-
out:
XXXXXXXX.mth File name of method
A-Meth.spec, esize: Identification for method report
and details of file size in bytes
Report MethSpc Method report (see above)
B-File Identification for binary file
XXXXXXXX.mth Method file (binary file)