Owner manual
5.5.3 CURVES
746 VA Trace Analyzer / 747 VA Stand
5-111
Softkeys Meaning
Scale
Axis scaling
Possibility for interactive modification of the parameters for the axis
scaling of the segment curves (display/plot parameters of the seg-
ments, see also section 5.6.2).
Unlinked
peaks
Non-assigned peaks
Display of all peak and wave voltages U.ver/mV found for the se-
lected segment but not linked to a substance.
Spike
elimin.
Spike elimination
Possibility for interactive modification of spike measured values.
This softkey is not available for curves recorded with
PSAMODE.
Parameter Range; Default value / Meaning
Segment-VR-
Curve
display only
Identification of selected curve
Segment Name of selected segment
VR Variation and replication number of the selected
segment
Curve
original Original curve
processed Processed curve
smoothed Smoothed and evaluated curve with baseline
derived Derived curve (1
st
derivative)
-backgnd Background curve (only with determinations with
BACKGND instructions, possible for all four curve
types)
13
14