Owner manual

5.5.3 CURVES
746 VA Trace Analyzer / 747 VA Stand
5-103
Substance-VR-Curve: XXXXXXXX-XX-smoothed
U.peak : XXXXXX mV dU.front: XXXXXX mV dU.rear: XXXXXX mV
I.peak : XXXXXX xA S.front: XXXXXX S.rear: XXXXXX
P.peak : XXXXXX xW
U.width: XXXXXX mV
Modify
column
Start
zoom
New
display
Reset
param.
Save
param.
Softkeys Meaning
Modify
column
Modify display/plot parameters column-by-column
Modification of the parameters for the axis scaling. For details of the
display/plot parameters, see section 5.6.3.
Modifications made here initiate a recalculation and curve display on
closure of the entry. These changes are temporary, they are canceled
with <Reset param.> or after exit with <ESC>.
The changes are permanently incorporated in the method only if they
are stored with <Save param.> before exiting.
Start
zoom
Select zoom range
This softkey starts the selection of the zoom region (curve section
which should be enlarged).
First, the cursor keys <è>, <ç>, <é> and <ê> must be used
to define the top left corner point of the zoom region (marked by a
cross) as follows:
Each time these keys are pressed, the corner point moves in the
selected direction by 5 pixels. A finer shift of 1 pixel can be
achieved when the <SHIFT> key is pressed at the same time. In
both cases, the cursor keys are repetitive, i.e. the longer the cur-
sor key is pressed, the further the corner point is shifted.
Uppr/lft
corner
Once the top left corner point is in the desired position, it must be
fixed by pressing the softkey <Uppr/lft corner> or the <ENTER>
key. Next, the bottom right corner point of the zoom region is also
defined using the cursor keys. This automatically opens a frame
encompassing the region to be enlarged.
9
Display / Plot
––––––––––––––––––––––––––
I.scale ........
I.begin ........A
I.end ........A
U.div ........V/cm
U.begin ........V
U.end ........V
8
ESC