Owner manual

5.5.3 CURVES
746 VA Trace Analyzer / 747 VA Stand
5-101
Parameter Range; Default value / Meaning
P.peak
Scope = whole Scope = f.half Scope = r.half
I
U
t.peak display only
Current evaluation quantity for PSA curves
Determined peak area in s.
t.peak is shown for curves recorded with the measurement mode
PSAMODE.
t.peak
t/U
U
U.width display only
Peak or wave width
Peak or wave width determined between the inflection points of the
peak or the derived wave in mV.
U.width is shown for curves recorded with one of the following
measurement modes:
DPMODE, SQWMODE, AC1MODE
DCTMODE
PSAMODE
CYCMODE with only 1 ramp
I
U
dI/dU
U
U.width
dU.front display only
Voltage difference to front base point
Absolute voltage difference determined between the base point at the
peak/wave front half and the peak voltage U.peak or the half-wave
potential U/2 in mV.