Owner manual
5.5.1 MONITORING
746 VA Trace Analyzer / 747 VA Stand
5-77
Parameter Range; Default value / Meaning
Start interval 0.0...99.9 h; 0.0 h
Time interval for automatic determination start
This parameter enables the automatic start of determinations at
regular time intervals for automatic series determinations with
Auto.batch = on (e.g. for online analysis).
The time interval set must be > the time needed for the whole deter-
mination, otherwise the next determination is started immediately.
Det.storage on, off; off
Automatic storage of determinations
The automatic storage of determinations in the determination memory
"DETERMINATIONS" can be switched on or off here.
on The determinations are automatically transferred to the
determination memory. If the memory is full, the pro-
cedure depends on the setting of the parameter Full
condition whereby either the oldest determinations
are deleted (Full condition = push out) or the de-
termination series is stopped (Full condition = stop,
see section 5.4.2).
off The determination data are not stored.
Determination
xxxx-xx-xx
xx:xx:xx
display only
Loaded determination
Name of the determination currently loaded in the working memory or
currently running. A name is given to each determination when it is
recorded and comprises the current time on method start and the
extension.dtm:
MMDDhhmm.dtm MM = month hh = hour
DD = day mm = minutes
The first, 8-digit part of the name can be modified on these dialog
pages after copying to the determination memory or the data card,
but the extension .dtm must be retained as it characterizes the
determination files.
Below the determination name, the exact time when the determination
was started is displayed.
Modified display only
Subsequent modification
Date and time when the determination loaded in the working memory
was subsequently modified for the first time (later modifications are
not displayed). The modification date is set only in the case of modifi-
cations on the following dialog pages:
COMMON VARIABLES Changes to Workg.com.var and the
associated Remarks
OP.SEQUENCE All changes
SEGMENTS All changes
MONITORING Modification of the parameters in the 1
st
sample data line, recalculation with soft-
key <Recalc. from Op.Sequence>).