5-54
Parameter Range; Default value / Meaning
Files display only
Number of files
Display of the number of stored determinations (maximum 256
determinations can be stored).
Available display only
Free storage space
Display of the free storage space in bytes.
File 8 characters;
Determination name
In copying a determination from the working memory to the determi-
nation list, the name MMDDhhmm.dtm is allocated automatically and
comprises the month (MM) – day (DD) – hours (hh) – minutes (mm) when
the determination was recorded and the extension .dtm. The first 8-
digit part of the determination name can be changed, however the
extension .dtm must be retained as it characterizes determination
files.
Secured determinations (*) must first be unsecured with <Unsecure>
before their name can be changed.
For the determination name all characters are allowed with the excep-
tion of the character ".".
Content 50 characters;
Content information for determination
In copying a determination from the working memory to the determi-
nation list, the three parameters Sample table, S1 and Method call
are entered automatically if they were available on the "MONITORING"
or "SAMPLES" page at the time of recording.
This 50-character content information can be changed at will and is
retained in direct copying to the data card or the RS interfaces.
However, if the determination is loaded in the working memory and
then copied from there to another destination, this content information
will be lost.
Secured methods (*) must first be unsecured with <Unsecure> before
their title can be changed.
Size display only
Memory size
Display of the memory requirements of the determination in bytes.
Full condition push out, stop; stop
Mode when memory full
Specification of what should happen when the automatic storage is
switched on (Det.storage = on on the "MONITORING" page), but the
determination memory is full:
push out Create space by deleting the oldest
determination(s)
stop Discontinue determination series
The mode can be switched with the softkey <Change full cd.>.