Owner manual

5.3 Set Up
746 VA Trace Analyzer / 747 VA Stand
5-40
Softkeys Meaning
To ensure the correct initializing of 700 Dosinos, the volume V.tube
in of the tubing from the supply bottle to the Dosino and V.tube out
of the tubing from the Dosino to the measuring vessel must be set.
To ensure the correct initializing of 685 Dosimats, the volumes V.tube
in and V.tube out must be set equal to the burette volume, so that
the burette cylinder is filled three times.
Fill
all
Fill all Dosimats/Dosinos
The burette cylinders of all Dosimats/Dosinos are filled at the set
Filling rate.
Should the filling take longer than 30 s, the message This operation
takes x.x minutes appears. The filling operation can then be started
with <r> or aborted with <a> to increase the filling rate. Once a filling
operation has been started, it can no longer be aborted.
After the filling, the initial volume V.init is automatically delivered to
the measuring vessel.
Parameter Range; Default value / Meaning
V.display display only
Dispensed volume
Display in mL of the volume dispensed with the Dosimat/Dosino.
V.burette display only
Burette volume
Display in mL of the burette volume of the exchange or dosing unit
mounted on the Dosimat/Dosino.
Type display only
Dosimat type
Display of the attached Dosimat type:
685 685 Dosimat
700 700 Dosino
Position dosing, exchange; dosing
Position of the Dosimat/Dosino
Display of the current position of the Dosimat/Dosino:
dosing Dispensing position
exchange Exchange position (exchange unit or dosing unit
can be changed)
If the Dosimat/Dosino is in the dosing position, it can be switched to
the exchange position by pressing <Exchange position>.
Gen.settings display only
General settings
Title for the general parameters of the Dosimats/Dosinos.