Owner manual
5.3.2 STAND
746 VA Trace Analyzer / 747 VA Stand
5-31
Softkeys Meaning
U.standby display only
Standby voltage at electrodes
Display of the polarization voltage set to the electrodes at the end of
the determination (U.standby is set with the ØMEAS command).
Rinse unit pos. exch., rinse; exch.
Position of the rinsing Dosinos
Display of the current position of the two rinsing Dosinos used for
rinsing the measuring vessel and siphoning off solution:
exch. Exchange position (dosing unit can be changed)
rinse Rinse position
To fill or empty the containers or to remove the dosing units from the
Dosinos, the Dosinos must be in the exch. position. If the Dosinos
are in the rinse position, they have to be switched to the exchange
position by pressing <Exchange position>.
Racks display only
Mounted racks
Display of the racks mounted on the 695 Autosampler:
S## Sample rack, e.g. S05 (6.2041.060 for 80 samples)
S10 (6.2041.070 for 40 samples)
S03 (6.2041.100 for 24 samples)
S12 (6.2041.110 for 32 samples)
A## Standard rack, e.g. A01 (6.2041.080 for 16 standards)
B## Standard rack, e.g. B01 (6.2041.090 for 10 standards)