Owner manual
5.3.2 STAND
746 VA Trace Analyzer / 747 VA Stand
5-21
5.3.2 STAND
STAND
Q
This dialog page is used for manual operation of the 747 VA
Stand(s) attached to the 746 VA Trace Analyzer and of the 695
Autosampler. It is accessible only in the instrument mode
* READY *.
747 VA STAND
Manual Address 1
Switches Settings States
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Pressure .... U.standby --- mV
Purging ....
Stirrer .... Rot.speed ..... /min Rinse unit pos. XXXXX
DME ....
HMDE .... Drop size .....
SMDE .... Drop size .....
RDE .... Rot.speed ..... /min
V.meas max ..... mL
V.rinse ..... mL
V.standby ..... mL
Pressure Purging Stirrer DME HMDE SMDE RDE >>
Settings
VA Stand
1
VA Stand
2
695 Auto
sampler
717 Smpl
Changer
Rinsing
Dosinos
Electr.
test
Default
values
>>
Switches
Softkeys Meaning
Pressure
Switch inert gas pressure on/off
The inert gas pressure needed for operation of the MME and for
purging is switched on or off. The current status is shown in the
Pressure parameter.
Even when the "STAND" page is quit, the inert gas pressure remains
switched on until it is switched off manually or automatically (by a
program start).
Purging
Switch inert gas purging on/off
Purging of the analyte solution is switched on or off. The current status
is shown in the Purging parameter.
Even when the "STAND" page is quit, purging remains switched on
until it is switched off manually or automatically (by a program start).
1