5.3 Set Up
746 VA Trace Analyzer / 747 VA Stand
5-20
Parameter Range; Default value / Meaning
Scan display only
Input control lines 1...8
Status display of the input control lines 1...8:
0 off, inactive (5 V)
1 on, active (0 V)
You will find further details on the control lines in section 3.13.
Set 0,1; 0
Output control lines 1...8
Here, the output control lines 1...8 can be set manually:
0 off, inactive (open)
1 on, active (0 V)
You will find further details on the control lines in section 3.13.
Ext. start on, off; off
External start via remote interface
Enable/disable external start via input control lines 1 and 2 of the
remote interface:
on External start/stop enabled:
Input control line 1 = on → <START>
Input control line 2 = on → <BREAK>
off External start/stop disabled
You will find further details on the control lines in section 3.13.
Printer display only
Printer at parallel interface
Title for the printer connected to the 746 VA Trace Analyzer.
Type Metrohm, IBM Propr., Epson 80*60, Epson 60*72, HP Deskjet,
HP Thinkjet, HP Thinkj.E; none
Printer type
Selection of the printer driver for the external printer attached to the
parallel interface (only for 746.0010 version without built-in printer).
Metrohm Built-in thermal printer
(only with 746.0020 version)
IBM Propr. IBM Proprinter and printers with IBM emulation
Epson 80*60 EPSON printers and printers with EPSON emu-
lation with 80 × 60 dpi graphics resolution (24-
pin dot matrix, ink jet and laser printers)
Epson 60*72 EPSON printers and printers with EPSON emu-
lation with 60 × 72 dpi graphics resolution (9-pin
dot matrix, ink jet and laser printers)
HP Deskjet HP printers (HP DeskJet, HP LaserJet) and
printers with HP PCL3 emulation
HP Thinkjet HP ThinkJet, HP QuietJet and compatibles
HP Thinkj.E HP ThinkJet in EPSON mode and compatibles
You will find further details on the attachment of printers and the
required settings in section 3.11.