Owner manual

4.6 Basic method for DPASV
746 VA Trace Analyzer / 747 VA Stand
4-65
4
Start the determination
Press the red <START> key.
The status display in the header of the dialog page "MONITORING"
switches to * BUSY *, the start time of the measurement is entered un-
der Determination and the determination is given a standard name,
also formed from the start time. Further, during the measurement the
measurement window is superimposed which shows the current pro-
gram line being executed and the remaining line time.
Shortly after the start of the determination, the screen shows the prompt
to add 10 mL sample to the measuring vessel.
5
Addition of sample
Remove stopper 2525 on 747 VA Stand and use a pipette to add 10 mL of
drinking water to the measuring vessel through the pipetting opening.
Press the <ENTER> key.
The screen shows the prompt to add 1 mL acetate buffer to the measur-
ing vessel.
6
Addition of acetate buffer
Use a pipette to add 1 mL acetate buffer (1 mol/L, pH 4.6) to the
measuring vessel through the pipetting opening. Then close the pipet-
ting opening using stopper 2525 and press the <ENTER> key.
7
Measurement of the sample solution
Measurement of the sample solution runs as follows (see also section
4.5.3):
Purging and stirring for 330 s.
Call-up of the segment ASV for measured value recording. The
substances present in the sample solution are enriched at the
hanging mercury drop with stirring at a measurement voltage of
–1150 mV and then determined anodically without stirring from
–1150 ... 150 mV. The single measurement is allocated the VR
number 00.
On completion of the measurement, the program is at an end. The status
display in the header of the dialog page "MONITORING" switches back
to * READY *, the instrument is ready for a new determination.
8
Printout of the segment curve
On completion of the program, the smoothed segment curve entered on
the "DOCUMENTATION" page is put out on the printer. The following ex-
ample shows a determination with drinking water as sample: