4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-64
5
Switch to dialog page "DOCUMENTATION"
• Press the yellow <DOCUMENTATION> key. The dialog page "DOCU-
MENTATION" appears for the automatic output of reports and curves.
The line entered here has the following meaning:
Curve Smth Sgmt:ASV VR:00 Printout of the smoothed segment curve
00 on the printer
• If you do not want an automatic printout in the following determination,
you can either delete this line with the softkey <Delete line> or clear
the output destination Printer for the line with the <CLEAR> key.
• Press the softkey <Comment>. The subpage of "DOCUMENTATION"
appears with the method comment describing, among other things, the
procedure used in the determination.
• Finally press the <ESC> key to return to the main page of "DOCUMEN-
TATION".
4.6.4 Performing the determination
To perform the determination, the following steps are necessary:
1
Switch to dialog page "MONITORING"
• Press the yellow <MONITORING> key. The dialog page "MONITORING"
whose header shows the loaded method ASV appears.
2
Switch the program run mode to determination
If the parameter Run mode is set to calibration, proceed as follows:
• Use the cursor keys to select the field Run mode.
• Press the space bar to enter the edit mode.
• Press the softkey <determ.> and confirm this entry with the <ENTER>
key.
3
Entry of the sample volume
Enter the sample volume as follows:
• Use the cursor keys to select the first field below the title Sample
size/S0.
• Press the space bar to enter the edit mode.
• Enter 10 mL and confirm this entry with the <ENTER> key.