4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-64
5
Switch to dialog page "DOCUMENTATION"
Press the yellow <DOCUMENTATION> key. The dialog page "DOCU-
MENTATION" appears for the automatic output of reports and curves.
The line entered here has the following meaning:
Curve Smth Sgmt:ASV VR:00 Printout of the smoothed segment curve
00 on the printer
If you do not want an automatic printout in the following determination,
you can either delete this line with the softkey <Delete line> or clear
the output destination Printer for the line with the <CLEAR> key.
Press the softkey <Comment>. The subpage of "DOCUMENTATION"
appears with the method comment describing, among other things, the
procedure used in the determination.
Finally press the <ESC> key to return to the main page of "DOCUMEN-
TATION".
4.6.4 Performing the determination
To perform the determination, the following steps are necessary:
1
Switch to dialog page "MONITORING"
Press the yellow <MONITORING> key. The dialog page "MONITORING"
whose header shows the loaded method ASV appears.
2
Switch the program run mode to determination
If the parameter Run mode is set to calibration, proceed as follows:
Use the cursor keys to select the field Run mode.
Press the space bar to enter the edit mode.
Press the softkey <determ.> and confirm this entry with the <ENTER>
key.
3
Entry of the sample volume
Enter the sample volume as follows:
Use the cursor keys to select the first field below the title Sample
size/S0.
Press the space bar to enter the edit mode.
Enter 10 mL and confirm this entry with the <ENTER> key.