Owner manual

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-60
6
Addition of acetate buffer
Use a pipette to add 1 mL acetate buffer (1 mol/L, pH 4.6) to the
measuring vessel through the pipetting opening. Then close the pipet-
ting opening using stopper 2525 and press the <ENTER> key.
7
Measurement of the sample solution
Measurement of the sample solution runs as follows (see also section
4.5.3):
Purging and stirring for 330 s.
Wait time 5 s to stabilize the analysis solution.
Call-up of the segment polaro for measured value recording.
The single measurement is allocated the VR number 00.
On completion of the measurement, the program is at an end. The status
display in the header of the dialog page "MONITORING" switches back
to * READY *, the instrument is ready for a new determination.
8
Printout of the segment curve
On completion of the program, the smoothed segment curve entered on
the "DOCUMENTATION" page is put out on the printer. The following ex-
ample shows a determination with drinking water as sample:
================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================
Method : POL User : Date: 1996-12-10
Mode : DPMODE Determ. : 12101420 Time: 14:20:12
Run : 6
Curve : smoothed Segment : polaro V R : 0 0