Owner manual

4.3 Lead determination with calibration curve
746 VA Trace Analyzer / 747 VA Stand
4-31
4.3.3 Method specifications of the method CALCRV.mth
In this section the 5 dialog pages of the group "Method Specifications" associated
with the loaded method CALCRV.mth will be explained in detail. If you are interested
only in a rapid lead measurement, you can proceed directly to section 4.3.4.
1
Switch to dialog page "OPERATION SEQUENCE"
Press the yellow <OP. SEQUENCE> key. The dialog page "OPERATION
SEQUENCE" appears with the main program for the lead determination:
Method: CALCRV .mth OPERATION SEQUENCE
Title : Pb determination with calibration curve
Instructions t/s Main parameters Auxiliary parameters
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
1 D SMPL>M V.fraction mL V.total L
2 C (CAL
3 C CSOLN>M addition into MEAS
4 DOS>M Soln.name Acetate V.add 1.000 mL
5 PURGE
6 STIR 180.0 Rot.speed 2000 /min
7 ØPURGE
8 (REP
9 STIR 5.0 Rot.speed 2000 /min
10 SEGMENT Segm.name DPASV
11 REP)2
12 C RINSE Cycles 1 V.rinse mL
13 C CAL)3
14 END
Modify
column
Modify
line
Delete
instr.
Insert
instr.
Method
labels
The following is a brief explanation of the main program sequence:
Line 1 Addition of sample: 10 mL
Line 2 13 Loop for recording the calibration curve (4 calibration solutions)
Line 3 Addition of calibration solution to measuring vessel: 10 mL
Line 4 Addition of acetate buffer to measuring vessel: 1 mL
Line 5 6 Purging and stirring of the analysis solution for 3 min
Line 7 Switching off the purging
Line 8 11 Loop for double replication of the segment
Line 9 Stirring of the analysis solution for 5 s
Line 10 Call-up of the segment DPASV for lead determination
Line 12 Resetting of the volume bookkeeping to 0 for solution change
(new calibration solution)
Line 14 End of main program
The sequence of the main program depends on the selected mode for
the program run Run mode on the "MONITORING" page:
In the program run mode calibration, all instructions marked with
D are ignored. 4 calibration solutions are measured in turn and the
coefficients of the calibration curve determined.
In the program run mode determination, on the other hand, all in-
structions marked with C are ignored. The sample is measured and
its mass concentration then determined using the calibration curve
recorded earlier.