Owner manual

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-26
2
Switch to dialog page "RESULTS/Evaluation"
Press the softkey <Eval.>. The dialog page "RESULTS" with the sub-
page "Evaluations" appears showing an overview of the substance and
peak evaluations of the determination loaded in the working storage:
Method : TESTPb RESULTS
Determ.: 03201614 Evaluations
Substance : Pb Comments
Mass conc.: 1.032 g/L Mass : 103.2 ug ––––––––––––––––––
MC.dev. : 0.011 g/L (1.05%) Add.mass : 100 ug
Cal.dev. : - V0.sample: 100 uL
VR U/mV I/nA I.mean Std.dev. I.delta Comments
––– ––––– –––––– –––––– ––––––– ––––––– ––––––––––––––––––
00 -355 -195.9 -194.7 1.569
01 -355 -195.3
02 -355 -192.9
10 -356 -385.2 -385.5 0.4942 -190.8
11 -356 -386.1
12 -356 -385.3
20 -357 -568.1 -568.5 1.968 -182.9
21 -357 -566.7
22 -357 -570.6
Select
subst.
Suspend/
activate
New
calcul.
>>
Calcul.
>>
Calibr.
You will find a brief explanation of the result data shown here in the
description of the full report in section 4.2.6.
On this subpage you have the possibility to ignore undesired peak
results (e.g. spikes) for subsequent calculations. Proceed as follows:
Use the cursor keys to select the desired result by its VR num-
ber.
Press the softkey <Suspend/activate>. In addition to the VR
number, the marking ! appears to show that this peak result will
be ignored in subsequent calculations (this marking can also be
canceled using the same softkey).
Press the softkey <New calcul.>. This initiates a new calculation
whose effects on the substance evaluation are immediately ap-
parent on this page.
3
Switch to dialog page "RESULTS/Calibration"
Press the softkey <Calibr.>. The dialog page "RESULTS" with the
subpage "Calibration" appears showing the results of the substance cali-
brations: