8 Troubleshooting
VA Database 2.0
91
8 Troubleshooting
8.1 Data transfer: What to do if ...
No data is received on the PC
1. Check cables and connections.
2. Make sure that the interface settings on the PC and the 746
VA Trace Analyzer are identical.
3. Check advanced interface settings on the PC (see Configure
Windows 95, page 3).
4. Press on the VA COMMUNICATION window.
5. If the problem continues, reduce the baud rate to 4800 or
2400 bits/s on the PC and the 746 VA Trace Analyzer.
The received data is not in the database
1. Check the destination database Associated database for the
COM port in the SERIAL INTERFACE SETTINGS window (see
Default directories, page 51).
2. Open this database and check the filtering criteria.
3. Make sure the an actual determination report Report ActDetm
has been sent form the 746 VA Trace Analyzer.
No data is received on the 746 VA Trace Analyzer
1. Check cables and connections.
2. Make sure that the interface settings on the PC and the 746
VA Trace Analyzer are identical.
3. Check advanced interface settings on the PC (see Configure
Windows 95, page 3).
4. Press on the VA COMMUNICATION window.
5. Set VA Trace Analyzer to receive mode by pressing <Copy
from> <RS232 Ifc.#>.