Manual
7 How to ...?
VA Database 2.0
71
Note: Each time a determination is sent to the database, a de-
termination file (*.dtm) and an ASCII determination report file (*.rpt)
will be stored automatically to the backup directory if the option
Save original files is enabled in the SERIAL INTERFACE SETTINGS
window (see Default directories, page 51).
Send methods to the backup directory (binary and
ASCII files)
1. If you want to observe the process of data transfer, click on
or MAIN WINDOW / Window / Communication to open the VA
COMMUNICATION window (the data transfer works also in the
background).
2. Load the method to be sent into the working storage of the
746 VA Trace Analyzer.
3. On dialog page "DOCUMENTATION" in the 746 VA Trace
Analyzer enter on a new line Report ActMeth in the COPY Re-
ports, Curves column and RSIfc.1 or RSIfc.2 in the TO destina-
tion column.
4. Activate the data transfer by moving the cursor to the de-
scribed line and pressing the softkey <Exec. line>.
Using this procedure both the binary method file (*.mth) and the
ASCII method report file (*.rpt) are transferred and stored in the
backup directory.
Note: If a method contains the line COPY Report ActMeth TO
RSIfc.# on the dialog page "DOCUMENTATION", the method files
are sent automatically to the backup directory after termination of
the analysis.
Send methods to the backup directory (only binary
files)
1. If you want to observe the process of data transfer, click on
or MAIN WINDOW / Window / Communication to open the VA
COMMUNICATION window (the data transfer works also in the
background).
2. Select the method(s) to be sent on the dialog page
"METHODS" or "DATA CARD" in the 746 VA Trace Analyzer.
3. Press the softkeys <Copy to> and <RS232 Ifc.1>/<RS232 Ifc.2>.
The binary method files (*.mth) will be transferred to the prede-
fined backup directory (see Default directories, page 51).