Manual

4 VA Database Explorer
VA Database 2.0
46
The segment curve window can also be opened
by drag and drop of the view item directly to an
empty space of the MAIN WINDOW.
Segment overlay curves
Overlay segment curves of several segments
can be opened by drag and drop of the view
item directly into an open segment curve window
in the MAIN WINDOW.
Note: Overlay curves can only be created if the units of the x and
y axes are identical (e.g. no overlay of VA and PSA curves).
4.5 Export from VA Explorer
Send determinations to VA Trace Analyzer from VA
Explorer
VA DATABASE EXPLORER / Determination / Send to COM#
Send the selected determination (highlighted
line) to the chosen COM port. The data transfer
is then carried out in the background. If you want
to see the progress of the data transfer, open the
VA COMMUNICATION window.
Note: Before sending determinations, the connection to the 746
VA Trace Analyzer has to be set active (see Opening/closing
COM ports, page 50) and the 746 VA Trace Analyzer has to be
set to receive mode by pressing the softkeys <Copy from> <RS232
Ifc.#> on the dialog pages "DETERMINATIONS" or "DATA CARD".
Export of determination files from VA Explorer
VA DATABASE EXPLORER / Determination / Save As...
Save the determination file (binary file with exten-
sion *.dtm) of the selected determination
(highlighted line) into the directory specified in
the SAVE DETERMINATION AS FILE window.