Manual
1 Introduction
VA Database 2.0
3
Configure Windows 95 for VA Database
It is recommended to install the COM port driver with FIFO-
Buffers. Proceed as follows:
1. <Start> → Settings → Control panel → System
2. In SYSTEM PROPERTIES window choose the tab Device Man-
ager.
3. In the tree expand Ports by clicking on + and mark COM1 or
COM2 depending to which port the 693/746 is connected.
4. Click on Properties, choose tab Port settings and click on
<Advanced>.
5. Change the settings as follows.
6. Click on <OK> in each window.
Configuration of 746 VA Trace Analyzer
On page "VA TRACE ANALYZER" the following settings are rec-
ommended:
Baud rate 9600
Data bits 8
Stop bit 1
Parity none
Handshake HW Short
Receive Mode on
Printer Type none
Printer Width 1.0
Remote keyboard off
When the configuration of the interface has to be changed, it is
important to save the new parameter by pressing the softkey
<Save Ifc.par.> (F7) before leaving the page.
Note: The COM port settings on both the 746 VA Trace Analyzer
and the PC have to be the same.