4.6 Determinations
■■■■■■■■■■■■■■■■■■■■■■
84
■■■■■■■■
743 Rancimat
Rinse the product thoroughly with distilled water.
Clean the measuring vessel covers (5-16), the PTFE tubes
(5-17) and the electrodes (5-18) with acetone or 2-propanol
and rinse with distilled water. Pre-clean with washing-up liquid in
the case of severe contamination.
Remove the protective ring (5-19) to be able to clean the electro-
des.
2
Clean the reaction vessels and accessories
Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.
Clean the reaction vessel covers with acetone or 2-propanol
and rinse with distilled water. Pre-clean with washing-up liquid in
the case of severe contamination.
Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.
Note
Replace the reaction vessel covers if they are not tightly on the
reaction vessel anymore or the material is obdurate.
3
Clean the tubings
Clean the silicone tubings (5-2) with acetone or 2-propanol and
rinse with distilled water. Pre-clean with washing-up liquid in the
case of severe contamination.
Then heat the silicone tubings in the drying oven for 2 hours at
80 °C.
4.6.6 Adjusting the method parameters during the determination
1
Open the dialog window Live parameters
In the Control window, click on the symbol
or the menu
item View Live parameters....
The following dialog window appears: