Owner's manual
Table Of Contents
- Table of contents
- Table of figures
- 1 Introduction
- 2 Overview of the instrument
- 3 Installation
- 4 Operation
- 4.1 Fundamentals of operation
- 4.2 Instrument and Program Settings
- 4.3 Program information
- 4.4 Calibration functions
- 4.5 Methods
- 4.6 Determinations
- 4.6.1 Preparing samples
- 4.6.2 Preparing the instrument and the accessories
- 4.6.3 Preparing the determination
- 4.6.4 Starting the determination
- 4.6.5 Cleaning the instrument and accessories
- 4.6.6 Adjusting the method parameters during the determination
- 4.6.7 Stopping the determination manually
- 4.6.8 Status of the live curve
- 4.7 Results
- 4.8 GLP functions
- 5 Handling and maintenance
- 6 Troubleshooting
- 7 Technical specifications
- 8 Conformity and warranty
- 9 Accessories
- Index

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
743 Rancimat
■■■■■■■■
83
4.6.4 Starting the determination
Before you can carry out the determination, the channels you intend to
use must be ready(see Chapter 4.6.8, page 86).
1
Start determination
■ Start of an individual determination: Click on the button
[Start] of the corresponding channel.
■ Start of all channels of Block A: Click on the button
.
■ Start of all channels of Block B: Click on the button .
After the determination has been started, the live curve is dis-
played for every channel:
As soon as a stop criterion defined in the method (see table 2, page
63) has been reached, the determination is ended:
The determination is automatically saved in the database
Repos.mrd. The channel is ready for starting a new determination.
2
Clean vessels and accessories
Note
Always clean the vessels and accessories directly after each
determination (see Chapter 4.6.5, page 83).
4.6.5 Cleaning the instrument and accessories
1
Clean the measuring vessels and accessories
■ Clean used measuring vessels after pouring off the measuring
solution with ethanol or 2-propanol (do not use acetone!). Pre-
clean with washing-up liquid in the case of severe contamination.