Owner's manual
Table Of Contents
- Table of contents
- Table of figures
- 1 Introduction
- 2 Overview of the instrument
- 3 Installation
- 4 Operation
- 4.1 Fundamentals of operation
- 4.2 Instrument and Program Settings
- 4.3 Program information
- 4.4 Calibration functions
- 4.5 Methods
- 4.6 Determinations
- 4.6.1 Preparing samples
- 4.6.2 Preparing the instrument and the accessories
- 4.6.3 Preparing the determination
- 4.6.4 Starting the determination
- 4.6.5 Cleaning the instrument and accessories
- 4.6.6 Adjusting the method parameters during the determination
- 4.6.7 Stopping the determination manually
- 4.6.8 Status of the live curve
- 4.7 Results
- 4.8 GLP functions
- 5 Handling and maintenance
- 6 Troubleshooting
- 7 Technical specifications
- 8 Conformity and warranty
- 9 Accessories
- Index

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
743 Rancimat
■■■■■■■■
71
AOCS Cd12b-92
Formula calculation according to standard AOCS Cd12b-92 with Tar-
get temperature = 110 °C.
AOM
Formula calculation according to standard AOM with Target temper-
ature = 97.8 °C.
free
Formula calculation with freely selectable Target temperature.
Formula parameters
Factor
Parameter for the calculation of the standard time, which indicates the
factor by which the time changes per 10 °C. This value can be determined
through the extrapolation (see Chapter 4.7.4, page 117) of results, which
have been recorded at different temperatures.
Range
0.001 ... 100
Target temperature
Target temperature for the calculation of the standard time.
Range
20 ... 300 °C
4.5.2.6 "Documentation" tab
This tab contains parameters for the automatic report output.