Owner's manual
Table Of Contents
- Table of contents
- Table of figures
- 1 Introduction
- 2 Overview of the instrument
- 3 Installation
- 4 Operation
- 4.1 Fundamentals of operation
- 4.2 Instrument and Program Settings
- 4.3 Program information
- 4.4 Calibration functions
- 4.5 Methods
- 4.6 Determinations
- 4.6.1 Preparing samples
- 4.6.2 Preparing the instrument and the accessories
- 4.6.3 Preparing the determination
- 4.6.4 Starting the determination
- 4.6.5 Cleaning the instrument and accessories
- 4.6.6 Adjusting the method parameters during the determination
- 4.6.7 Stopping the determination manually
- 4.6.8 Status of the live curve
- 4.7 Results
- 4.8 GLP functions
- 5 Handling and maintenance
- 6 Troubleshooting
- 7 Technical specifications
- 8 Conformity and warranty
- 9 Accessories
- Index

4.5 Methods
■■■■■■■■■■■■■■■■■■■■■■
70
■■■■■■■■
743 Rancimat
Formula 3 for stability time:
on | off (Default value: off)
If this option is activated, the automatically determined stability time will
be calculated with the entered values. The result will be saved under the
name entered in the field "R3".
4.5.2.5 "Standards" tab
This function can be used to automatically calculate the induction time for
a temperature deviating from the measuring temperature, the so-called
standard time. The settings here do not have any effect on the measuring
sequence or the evaluation. After finishing the measurement, a post-cal-
culation or recalculation is possible at any time in the determination over-
view (see Chapter 4.7.5, page 121).
Calculate standard time
on | off (Default value: off)
Activation/Deactivation of the calculation of the standard time according
to the following formula:
Standard time = Factor
(Temperature - Target temperature)/10
x induction time
Formula according to...
Selection
AOCS Cd12b-92 | AOM | free