Owner's manual
Table Of Contents
- Table of contents
- Table of figures
- 1 Introduction
- 2 Overview of the instrument
- 3 Installation
- 4 Operation
- 4.1 Fundamentals of operation
- 4.2 Instrument and Program Settings
- 4.3 Program information
- 4.4 Calibration functions
- 4.5 Methods
- 4.6 Determinations
- 4.6.1 Preparing samples
- 4.6.2 Preparing the instrument and the accessories
- 4.6.3 Preparing the determination
- 4.6.4 Starting the determination
- 4.6.5 Cleaning the instrument and accessories
- 4.6.6 Adjusting the method parameters during the determination
- 4.6.7 Stopping the determination manually
- 4.6.8 Status of the live curve
- 4.7 Results
- 4.8 GLP functions
- 5 Handling and maintenance
- 6 Troubleshooting
- 7 Technical specifications
- 8 Conformity and warranty
- 9 Accessories
- Index

4.5 Methods
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
743 Rancimat
Time Conductivity Endpoint(s) Analysis stops...
✔ ✔
after reaching the first criterion (conductivity or
endpoint). If the conductivity entered has been
reached, there is no further measuring until the
induction/stability time (endpoint) has been
reached.
✔ ✔ ✔
after reaching the first criterion. If the time or
conductivity entered has been reached, there is
no further measuring until the induction/stabil-
ity time (endpoint) has been reached.
Actions at end of determinations...
Shut off heater
on | off (Default value: off)
Automatic deactivation of the heating at the end of the determination.
Shut off gas flow
on | off (Default value: off)
Automatic deactivation of the gas flow at the end of the determination.
Caution
If the gas flow is automatically switched off, the reaction vessels must
be removed from the instrument within 30 to 60 min after the end of
the determination. If the vessels are not removed, a vacuum could
develop over the course of time, which would thus result in back suc-
tion of the measuring solution into the hot reaction vessel.
4.5.2.2 "Evaluation" tab
These settings serve for automatic evaluation of the curve. After finishing
the measurement, curves can also be reevaluated in the database (see
Chapter 4.7.3, page 112).