Owner's manual
Table Of Contents
- Table of contents
- Table of figures
- 1 Introduction
- 2 Overview of the instrument
- 3 Installation
- 4 Operation
- 4.1 Fundamentals of operation
- 4.2 Instrument and Program Settings
- 4.3 Program information
- 4.4 Calibration functions
- 4.5 Methods
- 4.6 Determinations
- 4.6.1 Preparing samples
- 4.6.2 Preparing the instrument and the accessories
- 4.6.3 Preparing the determination
- 4.6.4 Starting the determination
- 4.6.5 Cleaning the instrument and accessories
- 4.6.6 Adjusting the method parameters during the determination
- 4.6.7 Stopping the determination manually
- 4.6.8 Status of the live curve
- 4.7 Results
- 4.8 GLP functions
- 5 Handling and maintenance
- 6 Troubleshooting
- 7 Technical specifications
- 8 Conformity and warranty
- 9 Accessories
- Index

4.2 Instrument and Program Settings
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
743 Rancimat
Displaying the gas flow
Proceed as follows:
1
Open dialog window
■ Click on the menu item Tools ▶ Gas flow control ▶ Display
gas flow....
The following dialog window appears:
Note
This option is only available if no determination is running.
4.2.5 Recording the temperature
The temperature can be recorded at any time for each individual heating
block.
The program writes the recorded measured values into a text file contain-
ing the following values:
t(s) Time in seconds.
Block [°C] Temperature measured in the heating block.
Ext. Sensor [°C] Temperature measured with the external tem-
perature sensor.
The text file is saved under the name U#X-YYMMDD-hhmmss.txt in
the folder Log, # denoting the instrument number (1...4), X denoting the
block (A or B) and YYMMDD-hhmmss denoting the date and time when
the data logging started.