4.8 GLP functions
■■■■■■■■■■■■■■■■■■■■■■
154
■■■■■■■■
743 Rancimat
Click on the menu item Tools GLP test Gas flow... and
confirm with [Next >].
The following dialog window appears:
Enter the Gas flow and confirm with [Next >].
The following dialog window appears:
2
Test all channels
Measure the gas flow sequentially for all channels and enter the
values in the corresponding fields K1…K4 in L/h.
After finishing the measurement, confirm the entries with
[Next >].
The dialog window with the test results appears.