4 Operation
743 Rancimat
■■■■■■■■
151
2
Carry out GLP test with test resistor
■ Click on the menu item Tools ▶ GLP test ▶ Conductivity... and
confirm with [Next >].
The following dialog window appears:
■ Select the option Use test resistor.
■ Under Resistance, enter the value of the test resistor in Ohms
(10000 Ω for 6.2109.030) and confirm with [Next >].
■ Connect the test resistor to the electrode connector (3-2) indica-
ted in the dialog window and confirm with [Next >].
■ Continue with step 4.
3
Carry out GLP test with standard solution
■ Insert a measuring vessel, each filled with 60 mL standard solu-
tion, together with the measuring vessel cover at all measuring
position (see Chapter 4.6, page 75).