4 Operation
743 Rancimat
■■■■■■■■
147
2
Enter calibration points of the temperature sensor
■ Click on the menu item Tools ▶ GLP test ▶ Temperature...
and confirm with [Next >].
The following dialog window appears:
■ Enter a Sensor name for the temperature sensor.
■ Under Point 1…3, enter the Temperature and the Resistance
of the calibration point.
■ Confirm the entry with [Next >].
The following dialog window appears:
3
Enter test parameters
■ Under Channels used for test, select the required channels.
One channel per block has to be selected at least.
■ Under Test duration per channel, enter the required test dura-
tion and confirm with [Next >].